Introduction and aims
Ultrasound for catheter tunnel related complications is an acknowledged and quite standardized procedure. The ultrasound examination of intraperitoneal causes of catheter malfunction has been reported only for paediatric patients [1]. Even catheter malposition, a common cause of catheter malfuncation, can be evaluated by ultrasound [2]. The aim of the study was to evaluate the feasibility of colour Doppler for testing of malfunctioning peritoneal catheters in adults.
Methods
The ultrasound examination is performed in a supine position of the patient with partially filled peritoneal cavity. The peritoneal catheter remains connected to the dialysate double bag set. An ultrasound probe of at least 5 MHz is used to follow the intraperitoneal tract of the catheter. The catheter is flushed by means of gravity with dialysis fluid, and accompanied by colour doppler scans parallel to the catheter. Presence of colour flow jets is confirming patency of catheter sideholes.
Results
Five malfunctioning peritoneal catheter were evaluated by colour Doppler ultrasound. The catheters presented prevalently problems of dialysate outflow. Ultrasound reveiled permanent adhesion of ecogenic structures at the distal part of the catheter, whereas Doppler ultrasound absence of flow signals at the distal part of the catheter. Videolaparoscopy confirmed the ecographic suspect of occlusion of catheter sideholes by omental wrapping.
Conclusions
Ultrasound of the peritoneal catheter is a non invasive ambulatory bedside procedure making possible the examination of the intraperitoneal segment. Colour doppler is adding important information about patency of catheter sideholes. We consider colour doppler ultrasound helpful for functional testing of the peritoneal catheter in patients with dialysate flow problems.